MRDC 600 MHz Spectrometer

Magnet: Day 1 | Day 2 | Day 3 | Day 4 & 5 | Console: Day 1 | Day 2 & 3 | Day 4

Day 1: Delivery and Setup
Nanuc is proud to host the newest magnet at the University of Alberta, the Magnetic Resonance Diagnostic Center's (MRDC) 600MHz actively shielded system. Installation began on April 25th, 2005 by Ian Davies of Oxford Instruments. Equipment was unpacked and the magnet was roughed into position. The TMC anti-vibration legs were placed into their approximate positions and the shipping restraints were removed from beneath the cryostat. At the end of the day the outer vacuum container (OVC) was being purged with a turbo pump. Stay tuned for more installation updates.

The MRDC 600MHz in crate

The MRDC 600MHz in crate.

Anti-vibration TMC Legs

Anti-vibration TMC Legs

The cryostat suspended by A-frame.

Shipping restraints securing the helium bath.

Fitting on the legs.

4 large boxes of merchandise. 3 large crates of installation tools. The magnet crate and 4 250L dewars.

The Nanuc 500 MHz AS (right) and her new big sister the MRDC 600 MHz AS (left). Bruce is already designing the new steps for the 600 MHz!

   